148
5977 Series MSD Operation Manual
5
General Maintenance
To Replace a Filament in an EI Ion Source
Materials needed
•
Filament assembly (G2590-60053)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5-mm (8710-1570)
Procedure
1
Vent the MSD. See
2
Open the analyzer chamber. See
“To Open the Analyzer Chamber”
3
Remove the ion source. See
4
Remove the gold plated screw and washer for the filament(s).
WA R N I N G
The analyzer operates at high temperatures. Do not touch any part until you are sure
it is cool.