Operating in Electron Ionization (EI) Mode
3
5977 Series MSD Operation Manual
65
The largest influence on operating pressure in EI mode is the carrier gas
(column) flow.
lists typical pressures for various helium carrier gas
flows. These pressures are approximate and will vary from instrument to
instrument by as much as 30%.
If the pressure is consistently higher than those listed, refer to the online help
in the MassHunter Data Acquisition software for information on
troubleshooting air leaks and other vacuum problems.
In the
Instrument Control
view you can set up an MS Monitor for displaying this
vacuum reading. The vacuum can also be read on the LCP or from the Manual
Tune screen.
Table 13
Ion vacuum gauge reading
Column flow rate,
mL/min
Optional Gauge
reading, Torr
Performance
turbo
pump
Gauge reading, Torr
Diffusion
pump
Foreline reading, Torr
Diffusion
pump
0.5
3.18E–06
2.18E–05
34.7
0.7
4.42E–06
2.59E–05
39.4
1
6.26E–06
3.66E–05
52.86
1.2
7.33E–06
4.46E–05
60.866
2
1.24E–05
7.33E–05
91.784
3
1.86E–05
1.13E–04
125.76
4
2.48E–05
6
3.75E–05
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...