Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
103
The
Open
and
Closed
states are shown in the monitors as
1
and
0
respectively.
Figure 16
Reagent gas flow control module schematic
Table 18
Flow control module state diagram
Result
Gas A flow
Gas B flow
Purge
with Gas A
Purge
with Gas B
Pump out
flow module
Standby,
vented, or
EI mode
Gas A
Open
Closed
Open
Closed
Closed
Closed
Gas B
Closed
Open
Closed
Open
Closed
Closed
MFC
On
setpoint
On
setpoint
On
100%
On
100%
On
100%
Off
0%
Shutoff valve
Open
Open
Open
Open
Open
Closed
Gas A
(methane)
supply
Gas B
(other)
supply
Gas A
select valve
Gas B
select valve
Mass
flow
controller
Calibration
valve
Restrictor
Calibration
vial
Shutoff
valve
GC/MSD
interface
GC column
CI ion
source
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...