80
5977 Series MSD Operation Manual
3
Operating in Electron Ionization (EI) Mode
To Close the Analyzer Chamber
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
Procedure
1
Ensure all the internal analyzer electrical leads are correctly attached.
Wiring is the same for both the standard EI and CI ion sources. The
extractor EI ion source has an extra wire leading to the extractor lens.
The wiring is described in
and illustrated in
82. The term “Board” in the table refers to the
feedthrough board located next to the ion source.
Table 14
Source board wiring
Wire description
Attached to
Connects to
Green beaded (2)
Quad heater
Board, top left (HTR)
White with braided cover (2)
Quad sensor
Board, top (RTD)
White (2)
Board, center (FILAMENT-1)
Filament 1 (top)
Red (1)
Board, center left (REP)
Repeller
Black (2)
Board, center (FILAMENT-2)
Filament 2 (bottom)
Orange (1)
Board, top right (ION FOC)
Ion focus lens
Blue (1)
Board, top right (ENT LENS)
Entrance lens
White beaded (2)
Ion source heater
Board, bottom left (HTR)
White (2)
Ion source sensor
Board, bottom (RTD)
Brown (1)
Extractor lens
(Extractor EI ion source only)
Board, middle left
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...