82
5977 Series MSD Operation Manual
3
Operating in Electron Ionization (EI) Mode
Check the side plate O-ring. Ensure the O-ring has a
very
light coat of
Apiezon L high vacuum grease. If the O-ring is very dry, it may not seal
well. If the O-ring looks shiny, it has too much grease on it. Refer to the
5977 Series MSD Troubleshooting and Maintenance Manual for lubricating
instructions.
2
Close the side plate.
3
Reconnect the side board control cables.
4
Make sure the vent valve is closed.
5
Pump down the MSD. See
“To Pump Down the MSD in EI Mode”
6
If you are operating in CI mode, or if hydrogen or other flammable or toxic
substance is used for carrier gas,
gently
hand tighten the front side plate
thumbscrew.
Figure 14
Ion source wiring
FB = Feedthrough Board
Ion source
heater wires
Repeller
(red wire
from FB)
Filament 1
(white wires
from FB)
Ion focus lens
(orange wire
from FB)
Entrance lens
(blue wire
from FB)
Filament 2
(black wires
from FB)
Ion source
sensor wires
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...