158
5977 Series MSD Operation Manual
6
CI Maintenance
4
Trace the wires for the ion source heater and temperature sensor to the
feedthrough board. Use the pliers to pull on the metal connectors to
remove these four wires from the feedthrough board connections.
5
Remove the thumbscrews that hold the ion source in place.
6
Pull the ion source out of the source radiator.
Table 28
Standard CI ion source wires
Wire color
Connects to
Number of leads
Blue
Entrance lens
1
Orange
Ion focus
1
White
Filament 1 (top
filament)
2
Red
Repeller
1
Black
Filament 2 (bottom
filament)
2
C A U T I O N
Pull on the connectors, not on the wires.
WA R N I N G
The analyzer operates at high temperatures. Do not touch any part until you are sure
it is cool.