Figure 3-66
Channel Bit Rate Menu
3. Set the parameters for channel bit rate sweep.
Start
Specifies the starting point of the channel bit rate.
Stop
Specifies the stopping point of the channel bit rate.
Step
Specifies the step by which channel bit rate is incremented while
measuring TAA. When you specify the Step parameter together
with Start and Stop parameters, NOP will be computed and
updated automatically.
NOP
Specifies the number of points to measure TAA relative to the
change in channel bit rate. If NOP is specified together with Start
and Stop parameters, Step will be computed and updated
automatically.
4. Click the [Return] button to return to Channel Bit Rate Sweep menu.
5. Select the parameter you want to measure. In this example, [PW] is selected.