134
Chapter 3
VEE Measurement Program Operation
Measurement
5. Set the following parameters.
Channel Bit Rate
Specifies the speed at which data bits are
written on the media per second [bps].
Write Current
Specifies the current applied to the write head.
Sense Bias
Specifies the sense bias applied to the read
head.
Data Pattern
Specifies the data pattern to the write head.
Track #
Specifies the track number to be tested.
Squeeze Pos.
Specifies the distance between the test track
and the adjacent track (i.e, squeezing track).
See measurement definitions on Chapter 4.
OTRC Write Pos.
Specifies the distance where the write data
track is written. (not the same as squeezing
track). This distance is expressed as
percentage of the track pitch. See
measurement definitions on Chapter 4.
Meas. Type
Specifies the type of measurement sequence.
1)[Erase Write Meas.] means to perform a
band erase, write a specified data then
measure the parameter. 2)[Write Meas.] means
to write a data, then measure. 3) [Meas.]
means to measure only.
NOTE
You cannot change the
Function
parameter. It is fixed to TAA in the SNA Triple Track
Test.
The parameter values of Channel Bit Rate, Write Current and Sense Current that have been
set in the
Configuration
menu will automatically update similar parameters in this menu.
6. Click [Interpolation Setup] button to open the Interpolation Setup menu.
Figure 3-79
Interpolation Setup Menu
Use Best Fit Window
Specifies the use of best fit window. If
checked, the specified window size (%) will
be shifted between the high and low
thresholds in order to get the best correlation
to plot the interpolation lines. If unchecked,
the high and low threshold settings will be
used to plot the interpolation lines. Refer to
Figure 3-80 on page 135.
Window%
Specifies the size of the window to be used in
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com