184
Chapter 3
VEE Measurement Program Operation
Measurement
[Bathtub] Menu
Measurement
Procedure
1. Click on the [Bathtub] button from the BER Main menu to open the bathtub menu.
Exec. Optimize
Optimizes the setup parameters in the
[Optimize] menu for optimum measurement
results. Optimization is required whenever
you change the ‘User Data Rate’ and ‘Endec’
parameters.
Abort
Aborts unexpected long measurements.
2. Set the parameters, click on each button to enter a new value.
OTC Threshold
Specifies the off-track capability threshold.
The threshold limit can be set from 0 min. to 1
max.
Meas. Bytes
Specifies the number of bytes to be measured.
BER Pattern #
Specifies the pattern number to be used to
write the data pattern. The data pattern of the
selected pattern number must be specified in
the [Pattern Config.] menu in advance.
Average
Specifies the number of revolutions to get the
average result. Average count is equal to the
number of times a measurement sequence is
repeated.
Track #
Specifies the track number to be tested.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com