76
Chapter 3
VEE Measurement Program Operation
Measurement
Resolution Measurement (Parametric)
Overview
This section describes the Resolution measurement. Resolution is used to measure the TAA
of HF and LF data patterns. Resolution is measured by the parametric module (E5038A).
Measurement
Procedure
1. To open the Resolution menu, click on the [Resolution] button from the Parametric
frame of the Measure menu.
Figure 3-31
Resolution Menu
NOTE
Before you execute this measurement, you should set an appropriate input range or execute
the Auto Configuration to set the input range automatically.
2. Set the parameters, click on each button to enter a new value.
Channel Bit Rate
Specifies the speed at which data bits are
written/read on the media per second [bps].
Write Current
Specifies the current applied to the write head.
Sense Current
Specifies the current applied to the read head.
HF
Specifies the number of repetitive periods for
HF data pattern.
LF
Specifies the number of repetitive periods for
LF data pattern.
Track#
Specifies the track number to be tested.
Average
Specifies the number of revolutions to get the
average results.
NOTE
The parameter values of Channel Bit Rate, Write Current and Sense Current that have been