Switch on the System
This section provides the procedure when you turn on the system.
NOTE
Do not turn on the electronics system (VXI modules, oscilloscope, PC, monitor, etc.),
before you turn on the spinstand system. Doing so may cause some troubles or failures of
the spinstand system.
Step 1. Check the air pressure of the spinstand.
Step 2. Check the vacuum pressure of the spinstand.
Step 3. Switch on the spinstand.
Step 4. Switch on the VXI mainframe and other measuring instruments.
Step 5. Switch on the PC and its monitor.
Figure 2-2
Front Panel of the Spinstand
Check the air pressure and the vacuum pressure of the spinstand.
Make sure that the air pressure gauge and the vacuum pressure gauge readings are both
correct.
Table 2-1
Air Pressure and Vacuum Pressure
Air Pressure Range
0.67 MPa to 0.73 Mpa
(98 psi to 105 psi)