Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Frequency Subsystem ([:SOURce])
NOTE
To perform a frequency and amplitude sweep, you must also select LIST as the power mode.
See the
:MODE
command for selecting the list mode for an amplitude sweep.
*RST
CW
Key Entry
Freq
Freq Off
:FREQuency:MULTiplier
Supported
All Models
[:SOURce]:FREQuency:MULTiplier <value>
[:SOURce]:FREQuency:MULTiplier?
This command sets the multiplier for the signal generator carrier frequency. This displayed frequency
equals the actual frequency times the multiplier.
*RST
+1.00000
Range
Negative values: - 1000 to - 0.001
Positive values: 0.001 to 1000
Key Entry
Freq Multiplier
Remarks
For any multiplier other than one, the
MULT
indicator is shown in the frequency
area of the display.
:FREQuency:OFFSet
Supported
All Models
[:SOURce]:FREQuency:OFFSet <value><unit>
[:SOURce]:FREQuency:OFFSet?
This command sets the frequency offset.
The query of this command returns a value equal to the original output frequency times the
multiplier value, plus the frequency offset value. This displayed frequency equals the actual frequency
times the multiplier.
When an offset has been entered, the
OFFS
indicator is turned on in the frequency area of the
display.
The frequency offset state is turned on when any non–zero value is entered; entering zero will turn
it off. Refer to the
:FREQuency:OFFSet:STATe
command for setting the offset state independent of
entering offset values.
*RST
0.0000000000000E
00
Key Entry
Freq Offset
:FREQuency:OFFSet:STATe
Supported
All Models
Range
200 to 200 GHz
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...