228
Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Phase Modulation Subsystem–Option UNT ([:SOURce])
*RST
50 Ohm
:PM[1]|2:INTernal:DUAL:FUNCtion2:AMPLitude:PERCent
Supported
All Models with Option UNT and 303
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion2:AMPLitude:PERCent <value><unit>
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion2:AMPLitude:PERCent?
This command sets the amplitude of tone 2 of the internal dual function generator source as a
percent of the peak analog modulation amplitude. Tone 1 of the internal dual function generator
source will make up the remaining amplitude.
*RST
50.0
:PM[1]|2:INTernal:DUAL:FUNCtion2:POFFset
Supported
All Models with Option UNT and 303
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion2:POFFset <value><unit>
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion2:POFFset?
This command sets the phase offset in degrees or radians of tone 2 in relation to tone 1 of the
internal dual function generator source.
*RST
0.000 rad
:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:FREQuency
Supported
All Models with Option UNT and 303
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:FREQuency <value><unit>
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:FREQuency?
This command sets the frequency of tone 1 (default) or tone 2 of the internal dual function generator
source.
*RST
400.0 Hz
:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe
Supported
All Models with Option UNTand 303
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe SINE|TRIangle|SQUare|RAMP
Key Entry
Ext Impedance
50 Ohm
600 Ohm
1 MOhm
Range
0 to 100.0%
Key Entry
M Tone 2 Ampl % of Peak
Range
- 6.290 to 6.290 rad
- 360.4 to 360.4 deg
Key Entry
M Tone 2 Phase Offset
Range
100.0 mHz to 6.25 MHz
Key Entry
M Tone 1 Freq
M Tone 2 Freq
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...