Agilent X-Series Signal Generators SCPI Command Reference
227
Analog Modulation Commands
Phase Modulation Subsystem–Option UNT ([:SOURce])
Phase Modulation Subsystem–Option UNT ([:SOURce])
:PM[1]|2:BANDwidth|BWIDth
Supported
All Models with Option UNT
[:SOURce]:PM[1]|2:BANDwidth|BWIDth NORMal|HIGH
[:SOURce]:PM[1]|2:BANDwidth|BWIDth?
This command toggles between normal phase modulation and high- bandwidth phase modulation
mode.
*RST
NORM
Key Entry
FM
M Normal High BW
:PM[1]|2:EXTernal[1]|2:COUPling
Supported
All Models with Option UNT
[:SOURce]:PM[1]|2:EXTernal[1]|2:COUPling AC|DC
[:SOURce]:PM[1]|2:EXTernal[1]|2:COUPling?
This command sets the coupling for the phase modulation source through the selected external input
connector.
Use this command with the
EXTernal[1]|2:DC
command to remove the effects of DC and optimize the
DCFM calibration.
AC
This choice will only pass ac signal components.
DC
This choice will pass both ac and dc signal components.
*RST
DC
Key Entry
Ext Coupling DC AC
Remarks
This command does not change the currently active source or switch the current
modulation on or off. The modulating signal may be the sum of several signals,
either internal or external sources.
:PM[1]|2:EXTernal[1]|2:IMPedance
Supported
All Models with Option UNT
[:SOURce]:PM[1]|2:EXTernal[1]|2:IMPedance 50|600|1000000
[:SOURce]:PM[1]|2:EXTernal[1]|2:IMPedance?
This command sets the input impedance for the externally- applied phase- modulated input signal.
50
This choice selects 50 Ohm input impedance.
600
This choice selects 600 Ohm input impedance.
1000000
This choice selects 1 MOhm input impedance.
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...