Agilent X-Series Signal Generators SCPI Command Reference
377
Real–Time Commands
AWGN Real–Time Subsystem–Option 403 ([:SOURce]:RADio:AWGN:RT)
:IQ:MODulation:ATTen:AUTO
Supported
N5172B/82B with Option 403
[:SOURce]:RADio:AWGN:RT:IQ:MODulation:ATTen:AUTO ON|OFF|1|0
[:SOURce]:RADio:AWGN:RT:IQ:MODulation:ATTen:AUTO?
This command enables or disables the I/Q attenuation auto mode.
ON (1)
This choice enables the attenuation auto mode which optimizes the modulator
attenuation for the current conditions.
OFF (0)
This choice holds the attenuator at its current setting or at a selected value. Refer
to
“:IQ:MODulation:ATTen” on page 376
for setting the attenuation value.
*RST
1
Key Entry
Modulator Atten Manual Auto
:POWer:CONTrol
Supported
N5172B/82B with Option 403
[:SOURce]:RADio:AWGN:RT:POWer:CONTrol[:MODE] TOTal|NCHannel
[:SOURce]:RADio:AWGN:RT:POWer:CONTrol[:MODE]?
This command selects whether the instrument power is set by the displayed instrument power or the
channel noise power when the AWGN is turned on.
TOTal
This choice selects the displayed instrument power for control.
NCHannel
This choice selects the channel noise power for control. The channel noise power
is only settable from the front panel when the Power Control mode is set to
channel noise power.
*RST
TOTal
Key Entry
Power Control Mode
Total Nchannel
:POWer:NOISe:CHANnel
Supported
N5172B/82B with Option 403
[:SOURce]:RADio:AWGN:RT:POWer:NOISe:CHANnel <value>
[:SOURce]:RADio:AWGN:RT:POWer:NOISe:CHANnel?
This command sets the power within the channel bandwidth. The instrument power is changed in
relation to this setting if AWGN is turned on. The channel noise power is only settable from the front
panel when the Power Control mode is set to channel noise power.
The variable <value> is expressed in units of dBm.
*RST
Depends on model and options.
Key Entry
Channel Noise Power
:RATio
Supported
N5172B/82B with Option 403
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...