Agilent X-Series Signal Generators SCPI Command Reference
125
System Commands
Memory Subsystem (:MEMory)
Remarks
Refer to
“File Name Variables” on page 13
for information on the file name syntax.
:DATA:BIT:INSert
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:MEMory:DATA:BIT:INSert "<filename>",<bitOffset>,<bitCount>,<datablock>
This command inserts a specified number of bits at a specified bit offset in a bit file.
"<filename>"
This variable names the destination file and the directory path.
<
bitOffset
>
This number represents a value between 0 and a value that will fill all of the
available memory. This is where the <bitCount> bits of the data in the
<datablock> will be inserted. If the <bitOffset> is greater than the current file
length, then zero bits will fill the file from the current length in bits to the
<bitOffset>.
<
bitCount
>
This number represents the number of bits in the data block.
<datablock>
This parameter represents the data to be inserted into the file.
Refer to the
Programming Guide
for more information on downloading and using files.
Example
:MEM:DATA:BIT:INS "Test_Data",7,16,#12Qz
The preceding example inserts bit data at the 7th bit of the file,
Test_Data
. The table below
describes the command parameters.
Remarks
Refer to
“File Name Variables” on page 13
for information on the file name syntax.
2
Byte count
Qz
16 bits of data (ascii representation of bit data)
•
"Test_Data"
Test_Data
is the file name. The directory path is not
needed. The path
"/USER/BIT/"
is implied.
•
7
The offset in bits to insert the data in the data block
•
16
Number of bits in the data block
•
#12Qz
Data block
#
This character indicates the beginning of the data block
1
Number of digits in the byte count
2
Byte count
Qz
16 bits of data (ascii representation of bit data)
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...