Agilent X-Series Signal Generators SCPI Command Reference
149
System Commands
Route Subsystem (:ROUTe)
HARDware:DGENerator:OUTPut:DCS[:STATe]
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:ROUTe:HARDware:DGENerator:OUTPut:DCS[:STATe] ON|OFF|1|0
:ROUTe:HARDware:DGENerator:OUTPut:DCS[:STATe]?
This command is used to enable or disable the output DATA OUT, DATA CLK OUT, and SYM SYNC
OUT signals from the rear panel AUX I/O connector. Normally, these output signals should be enabled
(On). However, disabling these outputs will decrease the spurs that are sometimes present when
operating at high symbol rates.
*RST
1
Key Entry
DATA/CLK/SYNC Rear Outputs Off On
HARDware:DGENerator:OUTPut:DPOLarity
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:ROUTe:HARDware:DGENerator:OUTPut:DPOLarity POSitive|NEGative
:ROUTe:HARDware:DGENerator:OUTPut:DPOLarity?
This command configures the polarity of the TTL output signal at the DATA OUT connector. POSitive
refers to normal logic, while NEGative refers to inverted logic.
*RST
POS
Key Entry
Data Out Polarity Neg Pos
Remarks
This command performs the same function as
“HARDware:DGENerator:OPOLarity:DATA” on page 148
.
HARDware:DGENerator:OUTPut:SPOLarity
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:ROUTe:HARDware:DGENerator:OUTPut:SPOLarity POSitive|NEGative
:ROUTe:HARDware:DGENerator:OUTPut:SPOLarity?
This command configures the polarity of the TTL input signal at the SYMBOL SYNC connector.
POSitive refers to normal logic, while NEGative refers to inverted logic.
*RST
POS
Key Entry
Symbol Sync Out Polarity Neg Pos
LINE:PTRigger[1]|2:BNC:SOURce
Supported
N5172B with Option 653 or 655, N5182B with Option 656 or 657
:ROUTe:LINE:PTRigger[1]|2:BNC:SOURce BBTRigger[1]|BBTRigger2|EVENt[1]|PTRigger|NONE
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...