Agilent X-Series Signal Generators SCPI Command Reference
Digital Signal Interface Module Commands
Digital Subsystem—Option 003 and 004 ([:SOURce])
While there is a non–zero residual phase present in the signal, the DAC Over–Range Protection
feature will automatically prevent DAC overrange errors from occurring by scaling the signal down by
1/square root of 2.
Key Entry
Baseband Frequency Offset Phase Reset
:DIGital:DATA:INPut:FILTer:ALPHa
Supported
N5172B/82B with option 004
:DIGital:DATA:INPut:FILTer:ALPHa <value>
:DIGital:DATA:INPut:FILTer:ALPHa?
This command changes the Nyquist or root Nyquist filter’s alpha value.
The filter alpha value can be set to a minimum level (0), a maximum level (1), or in between by
using fractional numeric values (0.001–0.999).
Example
:DIG:DATA:INP:FILT:ALPH 1
The preceding example sets the filter alpha value to the maximum value of 1.
*RST
0.500
Range
0.000 to 1.000
Key Entry
Filter Alpha
:DIGital:DATA:INPut:FILTer:BBT
Supported
N5172B/82B with option 004
:DIGital:DATA:INPut:FILTer:BBT <value>
:DIGital:DATA:INPut:FILTer:BBT?
This command changes the bandwidth- multiplied- by- bit- time (BbT) filter parameter of the selected
Gaussian filter.
The filter BbT value can be set to the maximum level (1) or in between the minimum level (0.100)
and maximum level by using fractional numeric values (0.101–0.999).
Example
:DIG:DATA:INP:INP:FILT:BBT 1
The preceding example sets the filter BbT value to the maximum value of 1.
*RST
0.500
Range
0.000 to 1.000
Key Entry
Filter Alpha
Содержание 72B
Страница 28: ...Contents Agilent X Series Signal Generators SCPI Command Reference xxviii ...
Страница 48: ...18 Agilent X Series Signal Generators SCPI Command Reference SCPI Basics SCPI Basics ...
Страница 224: ...194 Agilent X Series Signal Generators SCPI Command Reference System Commands Unit Subsystem UNIT ...
Страница 458: ...428 Agilent X Series Signal Generators SCPI Command Reference Index ...