![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651063.webp)
DD+DIS033.07E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 1 / 27
06-2008 Type
5367/100
Agfa Company Confidential
5.4.3.2 Sharpness
Adjustment
(1) Check the DICOM conformance statement whether the modality sends the images
in page mode or image mode.
(2) Select the required smooth factor at the modality to define the sharpness of the
image. See
.
(3) When finished, print test images from the modality.
(4) Check the image quality together with the customer
(5) In case the image quality is not satisfying continue as described in the flowchart.
See Figure 15.
536401XK.CDR
sharp
too sharp
less sharp
299
199
101
200
148
143
245
235
251
138
Cubic High Res
Cubic Bell
Interpolation Smooth Factors (DICOM values)
299
200
251
245
235
Limits
Legend
Default
Useable values
Figure 18
NOTE:
Use
Cubic Bell
interpolation, when the modality sends the images in Page
Mode (images already composed for the film).
Use
Cubic High Res
interpolation, when the modality sends the images in
Image Mode (printer composes the different images for the film).
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......