![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 266](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651266.webp)
DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 14
06-2008 Type
5367/100
Agfa Company Confidential
(5) Print the Medical service test
image.
(6) Check the print stripes in the
medical image.
•
If the vertical stripes are
acceptable, continue with next
step.
•
If the vertical stripes are still
unacceptable, but improved
with respect to the first printed
medical image, repeat the
procedure.
•
If the vertical stripes are still
unacceptable and not
improved with respect to the
first printed medical image
replace the Thermal Print
head (refer to Section 3.5).
(7) Perform a Film calibration.
Figure 5: Medical service test image
NOTE:
With Print Head Calibration only macro TPH banding can be solved. Sharp and or
small (10 mm) discontinuities can not be solved by a Print Head Calibration.
2.4
Smart Profile Calibration
WHEN TO BE
PERFORMED
The smart profile calibration has to be performed in following cases:
•
During the troubleshooting procedure.
•
In certain cases during maintenance. Refer to section 9, Maintenance.
PURPOSE
Equalization of TH resistor drift in the area of 10" film black borders when printing in
mixed format conditions (mix of 10" and 14" print width).
A 14x17" film is required to perform this calibration.
66 densities have to be measured and to be entered in the browser screen.
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......