![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 396](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651396.webp)
DD+DIS032.07E
Maintenance
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 9 / 18
09-2007 Type
5367
Agfa Company Confidential
536631JN.cdr
(1) Insert the power cord at the printer
side; leave it disconnected at the
wall side.
(2) Put one probe at the ground
connector of the mains connector.
(3) Put the other probe at a few metallic
parts, the customer could touch.
The electrical resistance in this case
has to be
≤
0.3 Ohm.
Refer to Figure 15.
Figure 15
IMPORTANT:
If the electrical resistance of one measurement exceeds the limits (0.2 Ohm without
mains cable; 0.3 Ohm with mains cable) check the printer for loose grounding
connections!
3.9.2
Electrical Check according to National Regulations.
(1)
Perform the electrical check according to National Regulations.
NOTE:
The electrical checks that have to be performed in certain intervals (e.g. each 12
months) or after certain service actions (repair and/or maintenance) depend on
National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent regulations.
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......