DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 17
06-2008 Type
5367/100
Agfa Company Confidential
Verification:
(1)
In the browser screen select menu 'Service Engineer Tools - Enter Service
Command Session - 3 Print Image - 1 Testimage – upper or lower tray* -
STI2.ZIP': The STI2 test image will be printed.
(2)
From the same menu print the medical_image.zip*
*IMPORTANT: Select the tray with 14" width films.
(3)
Check both images at the light box for dark stripes in the area of the 10" film
edge:
•
On the STI2 test image dark stripes may be visible in the area of the 10"
film edge.
•
On the medical test image
no
dark stripes may be visible in the area of the
10" film edge.
Figure 9
Result:
The smart profile calibration is performed.
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......