![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 174](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651174.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 8 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC301007
: uppertraymodule.s1
Type
Failure
Context
W301007
Action
None
Reason1
S1 not connected
Cure1
Connect
S1
Reason2
Upper tray module defect
Cure2
Replace
Part
MOD01
HC301016
: upper.filmcal.required
Type
Failure
Context
W301016
Action
None
Reason
Upper tray needs calibration
Cure
Perform upper tray calibration
HC301018
: upperTag-unsupported
Type
Accidental
Context
Action
None
Reason
Wrong version in upper tray
Cure
Replace
filmpack
HC301019
: tag.history.insecure
Type
Defect
Context
E301019
Action
SeeManual
Reason
Tag history is corrupt
Cure
make Emergency sw install without preserving config
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......