![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 183](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651183.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 17 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC303021
: upperpickupmodule.m7.connection
Type
Failure
Context
W303021
Action
SeeManual
Reason1
Upper pickup motor not connected
Cure1
reconnect upper pickup motor
Reason2
Upper pickup module defect
Cure2
Replace
Part
MOD03
HC303022
: upperpickupmodule.m7.shortcircuit
Type
Defect
Context
E303022
Action
SeeManual
Reason
Upper pickup module defect
Cure
Replace
Part
MOD03
HC303023
: upperpickupmodule.s14
Type
Failure
Context
W303023
Action
None
Reason1
Upper pickup home sensor S14 malfunctioning
Cure1
check upper pickup home sensor S14
Reason2
Upper pickup module defect
Cure2
Replace
Part
MOD03
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......