![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 262](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651262.webp)
DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 10
06-2008 Type
5367/100
Agfa Company Confidential
2.2
CDM (Internal Densitometer) Calibration
NOTE:
•
Only following densitometers can be used for the calibration: McBeth TR-924,
Gretag D200-2, X-Rite 301, X-Rite 310, X-Rite 331, X-Rite 341.
•
The densitometer must be calibrated with the calibration wedge of the
densitometer manufacturer according to the calibration procedure of the
densitometer manufacturer.
WHEN TO BE
PERFORMED
•
The calibration of the internal densitometer CDM via an external densitometer is
necessary when the density reproduction check for a specific material is not OK
after a film calibration.
•
During the troubleshooting procedure.
PURPOSE
•
Adjust the CDM to another calibrated, external densitometer.
•
To compensate for CDM measurement drifts.
REQUIRED TOOLS:
One of following, calibrated densitometers: McBeth TR-924, Gretag D200-2, X-Rite
301, X-Rite 310, X-Rite 331, X-Rite 341.
INSTRUCTIONS
(1)
In the webpages - menu 'Service Engineer Tools - Enter Service Command
session' select '6. Calibration - 2. Internal densitometer calibration'.
(2)
Follow the instructions on the WebPages. Refer to Figure 3.
(3)
Enter the measured densities in the browser window.
(4)
Perform a film calibration. See 2.1.
(5)
Perform a density reproduction check. See 3.1.
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......