![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 401](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651401.webp)
DD+DIS032.07E
Maintenance
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 9 / 23
09-2007 Type
5367
Agfa Company Confidential
Select densitometer:
4
Dmax = 305
Square
Measured density
Target density Density diff. Check
100%
20
20
0
0
90%
46
48,5
-2,5
0
80%
74
77
-3
0
70%
102
105,5
-3,5
0
Density Reproduction =
OK
60%
130
134
-4
0
No MDM calibration needed for DT2B film
50%
158
162,5
-4,5
0
40%
186
191
-5
0
30%
215
219,5
-4,5
0
20%
244
248
-4
0
10%
270
276,5
-6,5
0
0%
300
305
-5
0
When entering the measured density values the decimal point appears in a different spot, e.g. measured value = 0.23, value to enter = 23.0
Density
Lower limit
Upper limit
0
0
0
30
-3
3
45
-4
4
60
-5
5
75
-6
6
90
-7
7
X-rite 310
105
-8
8
X-rite 331
120
-9
9
X-rite 341
135
-9
9
X-rite 301
150
-9
9
Mac Beth TR924
165
-9
9
Gretag D200-2
180
-10
10
195
-10
10
210
-10
10
225
-10
10
240
-10
10
255
-11
11
270
-11
11
285
-11
11
300
-11
11
315
-12
12
Density difference SMPTE image DT2B film DSAXYS
-15
-10
-5
0
5
10
15
0
50
100
150
200
250
300
350
Target Density
De
n
s
it
y
Dif
fe
re
nce
Lower limit
Upper limit
Density difference
X-rite 341
Figure 20
NOTE:
The measured density values have to be entered without decimal point
(e.g. measured value = .23, value to enter = 23)
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......