![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 256](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651256.webp)
DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 4
06-2008 Type
5367/100
Agfa Company Confidential
1
Adjustments
Following adjustments have to be performed at the printer:
Adjustment
To be done ....
Details see....
Horizontal
film
alignment
in case after TH replacement the blank left / right
frame on a SMPTE test image is < 1,5 mm for DT2B /
DT2C films or < 1 mm for DT 2 Mammo Films
1.1
Horizontal Film Alignment Procedure
WHEN TO BE
PERFORMED
The horizontal film alignment is necessary when:
•
The TPH has been replaced and
•
the blank frame at the left and right hand side on a SMPTE test image is
o
< 1,5 mm for DT2B / DT2C Films
o
< 1 mm for DT 2 Mammo Films.
PURPOSE
Centering the image on the film: Compensate for mechanical tolerances of the
thermal head mounting.
REQUIRED TOOLS:
Ruler to measure in the range up to 15 mm
REQUIRED TIME:
10 min.
NOTE:
The horizontal film alignment can be performed via browser user interface or the local
keypad. The required time and proceeding is identical.
It can be performed at any film size.
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......