![AGFA DRYSTAR AXYS Скачать руководство пользователя страница 220](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651220.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 54 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC318020
: devilmodule.tyndall.ram
Type
Defect
Context
18
Action
Replace
Reason1
RAM
defect
Cure1
Replace
Part
MOD18
Reason2
MOD12 cpu board (Thanos or Tyndall) defect
Cure2
Replace
Part
MOD12
HC312025
: devilmodule.tyndall.clock
Type
Defect
Context
12
Action
Replace
Reason
MOD12 cpu board (Thanos or Tyndall) defect
Cure
Replace
Part
MOD12
MODULE : powersupplymodule
Elabel : Yes
HC309000
: powersupplymodule.temperature
Type
Defect
Context
E309000
Action
SeeManual
Reason1
PSU too hot
Cure1
Switch off printer and let it cool down
Reason2
PSU
module
defect
Cure2
Replace
Part
MOD09
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......