DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 70 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC317002
: pumpandvalvesmodule.v2.connection
Type
Defect
Context
E317002
Action
SeeManual
Reason1
XS10 not connected
Cure1
Connect
XS10
Reason2
Pump and valves module defect
Cure2
Replace
Part
MOD17
HC317020
: pumpandvalvesmodule.m9.shortcircuit
Type
Defect
Context
E317020
Action
SeeManual
Reason
Pump and valves module defect
Cure
Replace
Part
MOD17
HC317021
: pumpandvalvesmodule.v1.shortcircuit
Type
Defect
Context
E317021
Action
SeeManual
Reason
Pump and valves module defect
Cure
Replace
Part
MOD17