DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 35 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC306011
: outputmodule.elabel.config
Type
Failure
Context
W306011
Action
None
Reason
Output elabel defect
Cure
Replace
Part
MOD06
HC306012
: outputmodule.cdm.home
Type
Accidental
Context
Action
None
Reason
CDM homing accidently failed
Cure
Self resolving accident
HC306013
: outputmodule.home
Type
Accidental
Context
Action
None
Reason1
Output homing accidently failed
Cure1
Self resolving accident
Reason2
Output module defect
Cure2
Replace
Part
MOD06
HC306014
: outputmodule.cdminstall
Type
Failure
Context
W306014
Action
None
Reason
CDM needs calibration
Cure
Calibrate
CDM
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......