OPTION 008 RF PROFILES AND COMPLEX SWEEP
Annex-B-12
RF
Info.
LF
Info.
Sweep
Sequence
LOCAL
Start
Sweep
Modify
Segments
Sweep
Type
Sweep
Mode
Sweep Sequence:
Sweep Status: WAITING FOR TRIGGER
Sweep Mode: INTERNAL SINGLE
Sweep Type: SEQUENCE SWEEP
C0859
0 - - - - - - - - -
Fig. B-12 Sweep sequence selection menu
Modifying segments
Sequenced sweeps are defined by a series of segments each of which has independent settings.
The segments can be constructed from the menu shown in Fig. B-13 called up by pressing the
[Modify Segments
] key.
Step
Size
Step
Time
Start
Freq.
RF
Level
Stop
Freq.
Segment
Number
LOCAL
Next
Segment
Previous
Segment
EXIT
Segment Number:
Start Freq :
Stop Freq :
RF Level :
Step Size :
Step Time:
C0856
0 -
10.0000
100.0000
100.0
20
+0.0
kHz
kHz
Hz
ms
Offset: +20.0 dB
Profile 1
dBm
Fig. B-13 Sweep segment editor
Up to 10 segments can be defined as Segments 0 to 9. If an existing segment is to be inspected (or
a segment similar to an existing one is required) the segment can be recalled by pressing
[Segment
Number
] followed by the segment number (0 to 9) and the [enter]
key. Pressing
[Next Segment
] or
[Previous Segment
] will increment or decrement through the segments. For each segment the
[Start Freq
.],
[Stop Freq
.],
[RF Level
],
[Step Size
] and
[Step Time
] keys can be used to define the
segment parameters.
Once the user has defined the required segments in a sweep pressing the
[EXIT]
key returns the
user to the
Sweep Sequence Selection Menu
shown in Fig. B-12.