GPIB OPERATION
4-11
RFLV
(continued)
RFLV?
Prepares message containing information on RF Level setting in the
following format:
:RFLV:UNITS <unit>;TYPE <type>;VALUE <nr2>;INC
<nr2>;<status>
where: <unit> is character program data defining the default RF level
units (DBM, DBV, DBMV, DBUV, V, MV or UV), <type> is
character program data indicating EMF or PD and <status> is a
program mnemonic indicating whether the RF output is ON or
OFF.
Examples:
:RFLV:UNITS DBM;VALUE -103.5;INC 2.0;ON
:RFLV:UNITS DBV;TYPE EMF;VALUE -83.2;INC 0.5;ON
RFLV:OFFS?
Prepares message containing information on RF Level offsets in the
following format:
:CFRQ:VALUE <nr2>;:RFLV:OFFS:VALUE
<nr2>;<status>;<activity>
where: <status> is a program mnemonic indicating whether the RF
offset is ON or OFF and <activity> is a program mnemonic
indicating whether the offset mode is enabled or disabled.
Example:
:CFRQ:VALUE 500000000.0;:RFLV:OFFS:VALUE -0.4;ON;ENABLE
RFLV
:HYST
[not used alone]
:ENABLE
Enable Extended Hysteresis mode
:DISABLE
Disable Extended Hysteresis mode
Data type :
None
Allowed suffices :
None
Default suffix :
None
:RFLV:HYST?
Responds with status as follows:
:RFLV:HYST:ENABLE
or
:RFLV:HYST:DISABLE
RFLV
:LIMIT
Set RF output level limit (short form)
:VALUE
Set RF output level limit.
Data type :
Decimal Numeric Program Data
Allowed suffices :
Any one of DBM, DBV, DBMV, DBUV, V, MV or UV
Default suffix :
dBm unless changed by UNITS command (see above).
:ENABLE
Enable RF output level limit
:DISABLE
Disable RF output level limit
:SAVE
Save RF output level limit to memory
Data type :
None
Allowed suffices :
None
Default suffix :
None
RFLV:LIMIT?
Prepares message containing information on RF level setting in the
following format:
:RFLV:UNITS <unit>;TYPE <type>;LIMIT:VALUE<nr2>;<status>
where: <unit> is character program data defining the default RF level
units (DBM, DBV, DBMV, DBUV, V, MV or UV), <type> is
character program data indicating EMF or PD and <status> is a
program mnemonic indicating whether the RF level limit is
enabled or disabled.
Examples:
:RFLV:UNITS DBM;LIMIT:VALUE -10.2;ENABLE
:RFLV:UNITS V;TYPE PD;LIMIT:VALUE 0.224;DISABLE