ANALOG MODULATION MODE
3-75
Calibration
Pressing
[Cal. Value]
brings the
Calibration Utilities Menu
to the display, see Fig. 3-73. This
menu shows when the last complete check was made and when the next calibration check is due. It
also shows the date on which the individual items were adjusted. It is possible to inspect the
calibration value of these items but calibration cannot be carried out unless the protection facility
is unlocked at Level 2. Full details regarding calibration can be found in the Service Manual.
Source/
Path
Digital
& Vector
FM
Selfcal
RF
Level
Int. Freq
Standard
Mod’n.
LOCAL
Calibration Utilities Menu
C1656
Last Complete Check : 1990-03-30
Next Cal. Check Due : 1992-03-30
Path / Src Gain Adjusted
Modulation Adjusted
FM Tracking Adjusted
RF Level Adjusted
Freq Standard Adjusted
: 1990-03-30
: 1990-03-30
: 1990-03-30
: 1990-03-30
: 1990-03-30
Fig. 3-73 Calibration utilities menu
Latch data
The latch data menu is intended for use as a diagnostic aid by allowing data to be sent to latches
within the instrument. For further information consult the Service Manual.
Elapsed time
The elapsed time facility displays the number of operating hours since the function was last reset.
Pressing
[Elapsed Time]
displays the number of operating hours and the date on which the
function was last set to zero. This facility can be used to assess the instrument's operational
reliability and utilisation.
Locking and unlocking
Press
[Lock & Unlock]
. When Level 1 and Level 2 are both locked, the menu displays three soft
keys:
Unlock Level 1
Unlock Level 2
Serial No. Set
Press
[Unlock Level 1]
and the message
Enter 4 Digit Password
: will appear on the display. Level
1 is unlocked by entering the 4 digits on the key pad and pressing [enter]. The menu will change
and two soft keys,
[Lock level 1]
and
[Lock Keyboard]
, will appear on the left-hand side. The
default password is 1234. If this password is not recognised by the instrument, the password has
been changed by your calibration/repair department personnel who should be consulted for further
information.
[Unlock Level 2]
is only used during servicing. Refer to the Service Manual for
details.