GPIB OPERATION
4-39
LF generator level
LFGL
Set LF Generator level (short form)
:VALUE
Set LF Generator level
Data type :
Decimal Numeric Program Data
Allowed suffices :
V, MV, UV, DBMV
Default suffix :
V
:INC
Set LF Generator level step
Data type :
Decimal Numeric Program Data
Allowed suffices :
DB
Default suffix :
DB
:UP
Go UP one step
:DN
Go DOWN one step
:RETN
Return to original setting
:XFER
Transfer current value to be the new setting
Data type :
None
Allowed suffices :
None
Default suffix :
None
:UNITS
Select default LF level units
Data type :
Character Program Data (DBM, DBV, DBMV, V, MV, or UV)
Allowed suffices :
None
Default suffix :
None
Examples:
LFGL:VALUE 75.6MV;INC 20DB
LFGL:UP
LFGL?
Prepares the message containing information on LF Generator Level
setting in the following format:
:LFGL:UNITS<unit>;VALUE<nr2>;INC<nr2>
where: <unit> is character programmed data defining the default LF
level units (DBM, DBV, DBMV, V, MV or UV).
Example:
:LFGL:UNITS MV;VALUE 125.8;INC 1.0