GPIB OPERATION
4-47
<YYYY-MM-DD>
where: <YYYY-MM-DD> is string program data representing the date
in ISO notation (year number, month number, day number).
Example:
"1990-04-01"
OPER?
Prepares message containing information on total operating hours in the
following format:
<nr2>
Example:
1453.0
ELAPSED?
Prepares message containing information on elapsed operating hours
since last reset in the following format:
<nr2>
Example:
454.5
ERROR?
Prepares message containing the number of the next error in the error
queue in the following format:
<nr1>
The numeric value returned is either that of the next error number or 0 if
the queue is empty or 255 if the queue is full
Example:
37
DEVTRG
Set Device Trigger Function (action on receipt of *TRG)
Data type :
Character Program Data (any one of: SEQT, FLSWP, SSSWP or VOID)
Allowed suffices :
None
Default suffix :
None
EXTTRG
Set External Trigger Function (action on Low signal being applied to
External Trigger Socket).
Data type :
Character Program Data (any one of: SEQT, FLSWP, SSSWP, MEMUP,
MEMDN or VOID)
Allowed suffices :
None
Default suffix :
None
Examples:
DEVTRG SEQT
EXTTRG MEMUP
KLOCK
Disables keyboard entry except RPP Reset and Go to Local
Data type :
None
Allowed suffices :
None
Default suffix :
None
KUNLOCK
Enables keyboard entry
Data type :
None
Allowed suffices :
None
Default suffix :
None