ACCEPTANCE TESTING
6-A-3
Table 6-A-1 Modulation oscillator frequencies
Modulation oscillator frequencies
(Hz)
10.00 279,620.20
139,810.10 500,000.00
Modulation oscillator distortion
C0219
LF
OUTPUT
UUT
Audio Analyzer
INPUT
➤
Fig. 6-A-2 Modulation oscillator distortion test set-up
(1)
Connect the test equipment as shown in Fig. 6-A-2.
(2)
Set the UUT such that the first modulation oscillator may be tested (refer to 'Modulation
oscillator frequencies' step (2) above).
(3)
Check that the distortion measured on the audio analyzer at the frequencies indicated in
Table 6-A-2 is less than 0.1%.
(4)
Set the UUT such that the second modulation oscillator may be tested (refer to
'Modulation oscillator frequencies' step (4) above).
(5)
Repeat (3) above.
Table 6-A-2 Distortion frequencies
Modulation oscillator frequencies
100 Hz
10 kHz
1 kHz
20 kHz