ACCEPTANCE TESTING
6-19
FM scale shape
(1) With the test equipment connected as in Fig. 6-9, set the UUT to
[Carrier Freq.]
31.64 MHz
[RF Level]
0 dBm
[FM Devn.]
224 kHz.
(2) Referring to Table 6-14, measure the FM on the modulation meter at the deviations indicated.
Table 6-14 FM scale shape deviations
Deviation (kHz)
224 256 289
228 261 293
233 265 298
238 270 302
242 275 307
247 279 312
252 284 316