ANALOG MODULATION MODE
3-68
Stopping the sweep
Press
[Stop Sweep]
. The sweep stops and the menu presents the opportunity to press:-
[Reset Sweep]
to change the sweep parameters, or
[Continue Sweep]
to continue the sweep, or
[Transfer]
to transfer the current value of the swept parameter as the last keyed in value in the
[SIG GEN] or [LF](
[LF Gen]
) mode, see Figs. 3-67 and 3-68. When the sweep is in the paused
state, the [
×
] and [
Ø
] keys can be used to step the parameter up or down. The sweep can then be
continued by pressing
[Continue Sweep].
LOCAL
Stop
Sweep
Reset
Sweep
Transfer
Level Start:
Level Stop:
Sweep Status: PAUSED AT +6.4 dBm
Sweep Mode: INTERNAL SINGLE
Sweep Type: RF LEVEL
Number of:
Steps
Step Time:
dBm
dBm
ms
-144.0
+13.0
1000
-----
50
C1901
Fig. 3-67 Sweep stopped
RF
Level
Carrier
Freq.
LOCAL
FM
ON/OFF
Select
Source
FM
Devn.
Source
Freq: F4
Low
Intermod.
Wideband
FM
AM
Φ
M
Carrier :
Freq.
RF Level :
dBm
Single Modulation Mode
Modulation ENABLED
MHz
Int Std: 10 MHz
FM:
kHz
Int F4:
1.0000 kHz
10.0
ON
626.407 6800
+6.4
ON
C1902
Fig. 3-68 RF level transferred
Aborting the sweep
Press
[Abort Sweep]
. The sweep is reset and the RF (or LF) signal is removed from the
appropriate output socket. The
Sweep Parameters Menu
as shown in Fig. 3-62 is displayed. If the
sweep is restarted with the signal disabled, the
[Abort Sweep]
key is not displayed thus giving the
user an indirect indication that no signal is being output from the instrument.