GPIB OPERATION
4-10
RF level
RFLV
Set RF output level (short form)
:VALUE
Set RF output level
Data type :
Decimal Numeric Program Data
Allowed suffices :
Any one of: DBM, DBV, DBMV, DBUV, V, MV or UV
Default suffix :
dBm unless changed by UNITS command
:INC
Set RF level step (dB)
Data type :
Decimal Numeric Program Data
Allowed suffices :
DB only
Default suffix :
DB
:UP
Go UP one step
:DN
Go DOWN one step
:RETN
Return to original setting
:XFER
Transfer current value to be the new setting
:ON
Turn RF output ON
:OFF
Turn RF output OFF
Data type :
None
Allowed suffices :
None
Default suffix :
None
:TYPE
Selects EMF or PD for voltage related units
Data type :
Character Program Data (EMF or PD)
Allowed suffices :
None
Default suffix :
None
:UNITS
Select default RF level units.
Data type :
Character Program Data (DBM, DBV, DBMV, DBUV, V, MV or UV)
Allowed suffices :
None
Default suffix :
None
Examples:
RFLV:VALUE -27.3DBM;ON
RFLV:TYPE PD;VALUE 1.23UV
:OFFS
[not used alone]
:VALUE
Set Offset of current band
Data type :
Decimal Numeric Program Data
Allowed suffices :
DB only
Default suffix :
DB
:ON
Turn ON offset of current band
:OFF
Turn OFF offset of current band
:ENABLE
Enable Offsets
:DISABLE
Disable Offsets
:SAVE
Store Offsets in memory
Data type :
None
Allowed suffices
:
None
Default suffix :
None
Example:
RFLV:OFFS:VALUE -0.2DB;ON;ENABLE