RK3000 High Frequency Power Supply
Operating Manual
Page 28 of 73
If the operating value is to be changed, simply press the numeric keys corresponding to
the new value (the old value will be overwritten) and press ENTER. If the present
operating value is acceptable, simply press ENTER without making any changes. The
Primary Current Limit display will now appear.
Primary Current Limit
The Primary Current Limit display is now shown.
The acceptable range for the Primary Current Limit is presented to the right of the word
“Range.” In this example the range is 1 to 500 amps.
The present operating value for the Primary Current Limit is presented to the right of the
word “Enter.” In this example the present operating value is 500 amps.
If the operating value is to be changed, simply press the numeric keys corresponding to
the new value (the old value will be overwritten) and press ENTER. If the present
operating value is acceptable, simply press ENTER without making any changed. The
Secondary Current Limit display will now appear.
Secondary Voltage Limit
The Secondary Voltage Limit display is now shown.
The acceptable range for the Secondary Voltage Limit is presented to the right of the
word “Range.” In this example, the present operating value is 99 kV (Kilovolts).
If the operating value is to be changed, simply press the numeric keys corresponding to
the new value (the old value will be overwritten) and press ENTER. If the present
operating value is acceptable, simply press ENTER without making any changes. The
display will now return to normal.
Secondary Current Limit
The Secondary Current Limit display is shown in below.
The acceptable range for the Secondary Current Limit is presented to the right of the
word “Range.” In this example the range is 1 to 1500 mA (milliamps).
RED -01
Pri Current Limit
Range: 1-500 Amperes
Enter: 500
RED -01
Sec. Voltage Limit
Range: 1-99 kV
Enter: 99
RED -01
Sec. Current Limit
Range: 1-1500 mA
Enter: 1500