Operation of the device
000000-1272-175-GA-GB-060613
89
The laser is intended for use with the focus in these three positions only.
1
Focus shift control
2
Magnification changer
Fig. 49
Focus shift control
Always turn the focus shift control until it clicks into position (an
audible click should be heard).
The focus shift of the VISULAS YAG III is designed for pulse energies of up to
about 1.5 mJ.
•
If you wish to work with higher energy levels (e.g. for fibrotic secondary
cataracts), you must increase the focus shift distance by defocusing:
–
In switch position ant., you must defocus in the anterior direction.
–
In switch position post., you must defocus in the posterior direction.
The degree of defocusing required depends on factors such as the
application and the specific conditions of the eye to be treated. It is
therefore not possible to give a general recommendation.
Summary of Contents for VISULAS YAG III
Page 1: ...VISULAS YAG III Documentation set...
Page 4: ......
Page 5: ...VISULAS YAG III User manual...
Page 6: ...000000 1272 175 GA GB 060613...
Page 122: ...Index 000000 1272 175 GA GB 060613 116 Switching on 56 Symbols 5 U Useful life 23...
Page 123: ......
Page 125: ......