Electromagnetic compatibility
000000-1272-175-GA-GB-060613
109
Guidance and manufacturer’s declaration - electromagnetic immunity
The VISULAS YAG III is intended for use in the electromagnetic environment specified below. The user of the
VISULAS YAG III should assure that it is used in such an environment
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be
that of typical commercial or
hospital environments.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
Mains power quality should be
that of typical commercial or
hospital environments.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5 % UT
(> 95 % dip in UT)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
< 5 % UT
(> 95 % dip in UT)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
Mains power quality should be
that of typical commercial or
hospital environments. If the user
of the VISULAS YAG III requires
continued operation during power
mains interruptions, it is
recommended that the
VISULAS YAG III be powered from
an uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.