Installation
000000-1272-175-GA-GB-060613
48
Installation of VISULAS Trion Combi
WARNING - GENERAL HAZARDS
This device may only be installed and operated for the first time by persons
who have been authorized by Carl Zeiss Meditec.
CAUTION - GENERAL HAZARDS
Fiber optics may not be tightly bent, kinked or incompletely fastened. The
device may otherwise be damaged, causing injury to the patient or user.
CAUTION - PROPERTY DAMAGE
Never touch the end of the optical fiber with your fingers and always
protect it from contamination.
Because of the laser’s high power output, any dirt may lead to damage to
the end surface of the fiber.
Screw in the optical fiber by hand only. Do not use tools!
Observe the user manuals of VISULAS Trion and the instrument
table.
•
Avoid impact and blows during shipment of the device.
•
Mount the system in a suitable location close to the physician and the slit
lamp.
•
Ensure that the VISULAS Trion is mounted on a flat surface.
•
Remove the key from the key switch to prevent the unit from being
inadvertently switched on during installation.
•
To install the VISULAS Trion on an instrument table with laser slit lamp, it
is best to roll the VISULAS Trion to the right of the instrument table
•
Now affix the laser slit lamp to the instrument table.
•
Push the VISULAS YAG III (laser console) onto the foot of the
VISULAS Trion from the back (Fig. 18).
For combined operation the foot switch must always be connected
to the VISULAS Trion, never to the VISULAS YAG III.