Installation
000000-1272-175-GA-GB-060613
54
•
Insert the plug of the ACCENTO eyepiece from below into the socket on
the microscope body of the slit lamp (3, Fig. 21). The plug and its lead
should be positioned towards the physician.
Data display in the ACCENTO eyepiece
When the ACCENTO eyepiece is used, in both treatment modes (STANDBY
and READY) treatment parameters are shown in the display of the slit lamp.
The energy value currently set is also shown at the top of the image field
Example:
0.9 mJ
Either the cumulative energy or the sum of the delivered pulses is
shown at the top of the image field, depending on which parameter has
been activated in the INFO mode.
Example:
∑
35.4 mJ
when the cumulative energy has been
activated
∑
28
when the sum of the delivered pulses has
been activated
When the focus shift display flashes (change of focus shift, change from
STANDBY treatment mode to READY treatment mode, counter reset)
the selected focus shift will appear briefly on the underside of the ACCENTO
eyepiece.
Example:
F POST
In combined operation (VISULAS YAG III Combi and
VISULAS Trion Combi) you can only switch between the display of
the cumulative energy and the sum of delivered pulses when the
coagulation laser (VISULAS 532s or VISULAS Trion) is switched off.
Switch on only the VISULAS YAG III and select the display as
setting will remain when the coagulation laser (VISULAS 532s or
VISULAS Trion) has been switched on.