Operation of the device
000000-1272-175-GA-GB-060613
85
READY
Fig. 46
Treatment READY mode
In Treatment READY mode, the aiming beam and laser warning lamp are
always on.
Parameters are set in the same way as for Treatment STANDBY mode.
•
When the trigger is activated, the laser beam will be triggered according
to the parameters set.
•
Touching the STANDBY/READY button returns you to
Treatment STANDBY mode.
•
Touching the Arrow button returns you to the main menu.
If, in Treatment READY mode, more than 5 minutes elapses
without the laser being triggered, the system returns to
Treatment STANDBY mode and emits a short audible signal.
Summary of Contents for VISULAS YAG III
Page 1: ...VISULAS YAG III Documentation set...
Page 4: ......
Page 5: ...VISULAS YAG III User manual...
Page 6: ...000000 1272 175 GA GB 060613...
Page 122: ...Index 000000 1272 175 GA GB 060613 116 Switching on 56 Symbols 5 U Useful life 23...
Page 123: ......
Page 125: ......