Maintenance and care
000000-1272-175-GA-GB-060613
96
Replace the halogen bulb (5, Fig. 51) as illustrated in Fig. 51.
•
The lamp housing can be removed by pressing down and simultaneously
pulling the lug (1, Fig. 51).
•
To unlock the lamp holder, turn the lever (8, Fig. 51).
•
Pull the socket plug (4, Fig. 51) from the bulb (5, Fig. 51).
•
Then remove the defective bulb (5, Fig. 51). Push a new bulb into the
opening, paying attention to the locating pins (7, Fig. 51). Do not touch
the bulb with the bare fingers.
•
Push the socket plug (4, Fig. 51) onto the pins of the bulb (5, Fig. 51).
•
Lock the lamp holder again with the lever (8, Fig. 51).
•
Insert the cover with holder (3, Fig. 51) into the slot (6, Fig. 51) of the
lamp housing and press it in until the retainers (2, Fig. 51) engage.
Summary of Contents for VISULAS YAG III
Page 1: ...VISULAS YAG III Documentation set...
Page 4: ......
Page 5: ...VISULAS YAG III User manual...
Page 6: ...000000 1272 175 GA GB 060613...
Page 122: ...Index 000000 1272 175 GA GB 060613 116 Switching on 56 Symbols 5 U Useful life 23...
Page 123: ......
Page 125: ......