Operation of the device
000000-1272-175-GA-GB-060613
69
Treatment READY mode
Touching the button STANDBY/READY in Treatment STANDBY mode
switches the system, after about three seconds, into
Treatment READY mode (Fig. 29).
Fig. 29
Treatment READY mode
In Treatment READY mode, the aiming beam and laser warning lamp are
always on.
Parameters are set in the same way as for Treatment STANDBY mode
(see page 64).
When the trigger is activated, the laser beam will be fired according to the
parameters set.
WARNING - RISK OF LASER RADIATION
Never trigger the laser if the aiming beam is not clearly visible in the target
area.
•
Touching the STANDBY/READY button returns you to
Treatment STANDBY mode.
If, in Treatment READY mode, more than 5 minutes elapses
without the laser being triggered, the system returns to
Treatment STANDBY mode.