Maintenance and care
000000-1272-175-GA-GB-060613
92
Error messages on YAG III
Error messages on VISULAS YAG III Combi/VISULAS Trion Combi
If, in YAG mode, an error occurs on the VISULAS YAG III, an error message
will be displayed on the control panel.
At the same time, an error message with a specific error code will be
displayed on the YAG console.
Correct the malfunction as instructed in the VISULAS 532s or VISULAS Trion
user manual. Then switch the VISULAS 532s or VISULAS Trion off and on
again.
Faults on the VISULAS 532s/VISULAS Trion are indicated as system
information on the control unit display. The user is also alerted by short
audible signals.
Message
Remedy
LASER STOP is activated; please
unlock.
Unlock the LASER STOP button.
Please release actuator!
Release foot or hand switch.
Door interlock is active, please
close door.
Shut the door or check that the shorting plug is correctly plugged into
the yellow socket.
Excess temperature, please wait!
The internal monitoring systems have detected an excess temperature.
Please wait a moment.
If this fault occurs frequently, ensure that the instrument’s ventilation
slots are not covered and that the ambient temperature falls below
35 °C.
Slit lamp not connected!
Check that the slit lamp is properly connected.
No foot switch connected!
Connect the foot switch or operate the instrument using the hand
switch. Select the required triggering mode (see
page 66 and following).
In the STANDBY and READY
modes, demo is displayed beside
the info button. The laser cannot
be triggered.
Switch the instrument off and then on again. While the progress bar is
active during the system test (see
ZEISS logo on the screen, then press the rotary knob/button.
Message
Remedy
In the STANDBY and READY
modes, demo is displayed beside
the aiming beam button. The
laser cannot be triggered.
Switch the VISULAS YAG III off and then on again. While the progress
bar is active during the system test on YAG III (see
page 62), touch the ZEISS logo on the screen, then press the rotary
knob/button.
Swivel lamp casing out of
position.
Swivel the lamp casing out of position. Check that all the connectors on
the rear of both consoles are inserted properly.