Version 9.0
G-30-1781-en
Page 131
S100 / OPMI pico
Operation
The main and backup lamp are intact in the halogen light source.
Main and backup lamp are intact in the xenon light source.
If the maximum service life of 500 hours of the ,main lamp has been
exceeded and the replacement lamp is being used, you should have
a replacement lamp module available for replacement purposes as
a precaution (see Page 226). For the order number, please see the
chapter "Device data".
With the LED light source, the amber LED is not illuminated.
It is possible to switch on or off the halogen, xenon or LED light source
by moving up and down the suspension arm.
The light source is turned on within the preset working area.
The light source is turned off in the upper resting position of the
suspension arm.
WARNING
Function deterioration of the light source
With increasing service life of the light source, the actual light intensity
decreases (this is a normal system property).
•
Before every use, ensure that the available light quantity is sufficient.
•
Have a backup light source (e.g. surgical lamp) available.
For halogen light source:
•
If necessary, exchange the lamp (seePage 224).
After changing the lamp, check the brightness setting.
For xenon light source:
•
Switch to the backup lamp.
If you are already working with the backup lamp, be sure to replace the
slide-in lamp module (see Page 226). Check the set brightness after the
lamp exchange.
WARNING
Precautions in case of device failure
•
We recommend you to take adequate precautions, depending on the
application, to enable the surgical procedure or treatment to be finished
without using this microscope, if required.
Summary of Contents for S100
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Page 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Page 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Page 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Page 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Page 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Page 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Page 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Page 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Page 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...