S100 / OPMI pico
Safety measures
Safety features
Halogen light source
1
Flap
The flap is used as mechanic indicator of the operating status of the
halogen light source.
–
When the flap is closed, the main lamp is operative.
–
When the flap is open, the main lamp has failed. The backup lamp
is on.
2
Switching to the backup lamp
The lamp housing contains a backup lamp. If the first lamp fails, you can
switch on the backup lamp by pressing this button.
•
Replace the defective lamp at the first opportunity (see Page 224).
Fig. 1: Safety device for the
halogen light source
2
1
Summary of Contents for S100
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Page 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Page 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Page 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Page 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Page 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Page 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Page 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Page 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Page 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...